Medical Device Apps: An Introduction to Regulatory Affairs for Developers

Ulrika SH Simonsson, Department of Pharmaceutical Biosciences, Uppsala University, Box 591, Uppsala, 75124, Sweden, Phone: 46 18 471 4000, Email: es.uu.oibmraf@nossnomiS.akirlU .

Lina Keutzer

1 Department of Pharmaceutical Biosciences, Uppsala University, Uppsala, Sweden

Find articles by Lina Keutzer

Ulrika SH Simonsson

1 Department of Pharmaceutical Biosciences, Uppsala University, Uppsala, Sweden

Find articles by Ulrika SH Simonsson Corresponding author. # Contributed equally. Corresponding Author: Ulrika SH Simonsson es.uu.oibmraf@nossnomiS.akirlU Received 2019 Dec 20; Revisions requested 2020 Feb 22; Revised 2020 Feb 28; Accepted 2020 Mar 28.

Copyright ©Lina Keutzer, Ulrika SH Simonsson. Originally published in JMIR mHealth and uHealth (http://mhealth.jmir.org), 26.06.2020.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mHealth and uHealth, is properly cited. The complete bibliographic information, a link to the original publication on http://mhealth.jmir.org/, as well as this copyright and license information must be included.

Abstract

The Poly Implant Prothèse (PIP) scandal in France prompted a revision of the regulations regarding the marketing of medical devices. The new Medical Device Regulation (MDR [EU]) 2017/745 was developed and entered into force on May 25, 2017. After a transition period of 3 years, the regulations must be implemented in all EU and European Economic Area member states. The implementation of this regulation bears many changes for medical device development and marketing, including medical device software and mobile apps. Medical device development and marketing is a complex process by which manufacturers must keep many regulatory requirements and obligations in mind. The objective of this paper is to provide an introduction and overview of regulatory affairs for manufacturers that are new to the field of medical device software and apps with a specific focus on the new MDR, accompanying harmonized standards, and guidance documents from the European Commission. This work provides a concise overview of the qualification and classification of medical device software and apps, conformity assessment routes, technical documentation, clinical evaluation, the involvement of notified bodies, and the unique device identifier. Compared to the previous Medical Device Directive (MDD) 93/42/EEC, the MDR provides greater detail about the requirements for software qualification and classification. In particular, rule 11 sets specific rules for the classification of medical device software and will be described in this paper. In comparison to the previous MDD, the MDR is more stringent, especially regarding the classification of health apps and software. The implementation of the MDR in May 2020 and its interpretation by the authorities will demonstrate how app and software manufacturers as well as patients will be affected by the regulation.

Keywords: MDR, medical device regulation, medical devices, medical device software, mHealth, eHealth, mobile apps, smartphone apps

Introduction

Due to safety issues in the field of medical devices, and especially after the Poly Implant Prothése (PIP) scandal in France, the Medical Device Directive (MDD) 93/42/EEC [1] was revised and replaced with the new Medical Device Regulation (MDR [EU]) 2017/745 [2,3]. The MDR entered into force on May 25, 2017 and must be implemented within the European Union (EU) and European Economic Area (EEA) states after 3 years, by May 26, 2020 [2]. Since the MDR is a regulation, it is immediately enforceable as law in all member states after its implementation date. This contrasts with the previous MDD, which was a directive that member states transpose into national law within a set timeframe [4].

Unlike the US Food & Drug Administration, which regulates foods, medicines, and medical devices, the European Medicines Agency (EMA) regulates only drugs. There is no regulatory body like the EMA for the review and approval of medical devices. Manufacturers themselves declare conformity of their devices with the European legislations and regulations and affix a CE (Communauté europeénne) mark (Article 10 and 20 MDR). Products that bear a CE mark can then be marketed within the EU/EEA (Articles 2 and 10 MDR). Affixing the CE mark to a product is only legal after a conformity assessment has been performed (Article 20 MDR). Depending on the class of the device, a notified body must be involved in this process (Introduction [60] and Articles 52-53 MDR). For certain highly critical or novel products, an additional examination by so called “expert panels” is mandatory (Introduction [56] and Article 106 MDR).

Independent of the risk class, technical documentation (TD) must be compiled to allow an assessment of whether the general safety and performance requirements set by the MDR are met (Annex II MDR). With the exception of class l devices, the notified bodies then inspect the manufacturer’s Quality Management System (QMS) and technical documentation and subsequently issue the required Annex certificates (Annex XII MDR). These are prerequisites for declaring conformity and affixing the CE mark (Article 10[6] MDR). For guidance on how to perform the steps required for CE marking, including risk management, technical documentation, and QMS, and to prove regulatory compliance, manufacturers are advised to work according to harmonized standards and common specifications (Articles 8-9 and Annex II [4c] MDR). When the MDR came into force in 2017, there was no associated harmonized standard or common specification. According to the European Commission, these will be implemented soon [5].

When developers of software or mobile apps claim that their product has a medical purpose, it becomes a medical device and must bear a CE mark (Article 2[1] MDR). This paper describes the process of placing mobile apps and software on the market as medical devices ( Figure 1 ) and serves as an introduction to regulatory affairs for app and software developers.

An external file that holds a picture, illustration, etc. Object name is mhealth_v8i6e17567_fig1.jpg

Important stages of medical device development.

Elements of the Medical Device Regulation

The primary elements of the new Medical Device Regulation (MDR [EU]) 2017/745 and the accompanying harmonized standards and guidance documents provided by the European Commission are described below.

Qualification of Mobile Apps and Software: What Constitutes a Medical Device?

Medical Device Software

Mobile apps and software that are independent of any device and are not intended to be used as an accessory to a medical device are referred to as Medical Device Software (MDSW) or standalone software [6] and must be qualified and classified in their own right (Annex VIII [3.3] MDR).

Intended Purpose

The first, essential question an app or software developer must answer is whether the product is a medical device or not. Software qualifies as a medical device if the developer’s stated purpose of the software meets the definition of a medical device in Article 2[1] of the MDR (Textbox 1).

Medical device definition [Article 2(1) MDR].

“Medical device” means any instrument, apparatus, appliance, software, implant, reagent, material or other article intended by the manufacturer to be used, alone or in combination, for human beings for one or more of the following specific medical purposes:

diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease, diagnosis, monitoring, treatment, alleviation of, or compensation for, an injury or disability,

investigation, replacement or modification of the anatomy or of a physiological or pathological process or state,

providing information by means of in vitro examination of specimens derived from the human body, including organ, blood and tissue donations, and which does not achieve its principal intended action by pharmacological, immunological or metabolic means, in or on the human body, but which may be assisted in its function by such means. The following products shall also be deemed to be medical devices:

devices for the control or support of conception;

products specifically intended for the cleaning, disinfection or sterilisation of devices as referred to in Article 1(4) and of those referred to in the first paragraph of this point.

The decision of whether a software product qualifies as a medical device is made by the developer or, using the terminology of the MDR, the manufacturer (Article 2[30] MDR). If the manufacturer states that the system can be used for a medical purpose, it must be CE marked (Articles 10 and 20 MDR). Whether a product qualifies as a medical device is determined by the intended use, as stated by the manufacturer and the mechanism of action of the product, not the design or user [6]. Furthermore, the description of the intended purpose must include a statement of benefit for the patient, otherwise it cannot be marketed as a medical device (Articles 61-62, Annexes XIV and XV MDR).

Key Characteristics for Qualification as a Medical Device

If a mobile app or software performs an action on data beyond storage, archiving, communication, or simple search; the performed action is for medical purposes; and the performed action is for the benefit of an individual patient, it most likely qualifies as a medical device and is subject to the MDR [6]. For further guidance on qualification of standalone software, see “MEDDEV 2.1/6 Guidelines on the qualification and classification of standalone software” [7].

Classification of Standalone Software Including Apps

The MDR defines four risk classes: I, IIa, IIb, and III ( Table 1 ). For classification of software, rule 11 has been included in the regulation (Textbox 2) (Annex VIII MDR).

Table 1

Summary of differences between risk classes.

ClassDocumentationNotified body involved?QMS a CertificatesClinical investigation
I (low risk)Manufacturer must compile the technical documentation and self-declare conformityNoYesNoNot mandatory. May be required depending on the outcome of the clinical evaluation
IIa (low-medium risk)Manufacturer must draw up the technical documentation and apply to a European Notified BodyYesYes, certifiedYes (Annex IX certificate, QMS certificate)Not mandatory. May be required depending on the outcome of the clinical evaluation
IIb (medium-high risk)Manufacturer must draw up the technical documentation and apply to a European Notified BodyYesYes, certifiedYes (Annex IX certificate, QMS certificate)Not mandatory. May be required depending on the outcome of the clinical evaluation
III (high risk)Manufacturer must draw up the technical documentation and apply to a European Notified BodyYes, expert panelYes, certifiedYes (Annex IX certificate, QMS certificate)Mandatory

a QMS: Quality Management System.

Rule 11, Annex VIII MDR.

Software intended to provide information which is used to take decisions with diagnosis or therapeutic purposes is classified as class IIa, except if such decisions have an impact that may cause:

Death or an irreversible deterioration of a person's state of health, in which case it is in class III; or

Serious deterioration of a person's state of health or a surgical intervention, in which case it is classified as class IIb.

Software intended to monitor physiological processes is classified as class IIa, except if it is intended for monitoring of vital physiological parameters, where the nature of variations of those parameters is such that it could result in immediate danger to the patient, in which case it is classified as class IIb. All other software are classified as class I.

This rule implies that many apps might have to be classified as class IIa, IIb, or III in the future, while under the former Medical Device Directive (MDD) 93/42/EEC most standalone software including apps were classified as class l or not designated as medical devices at all [1]. Interpreting rule 11 on its own indicates that, for example, software used to calculate dosages of drugs with high toxicity, suggesting a diagnosis, or aiding with therapy or radiation planning will fall within class III, since a mistake might cause death. If it is highly unlikely that death could be caused by an error, it could fall within class IIb, which is defined as a device for which “…a mistake can cause serious deterioration of a person’s state of health. ” (Annex VIII MDR). Medical device software may only fall under class IIa if a mistake cannot be anticipated to cause serious deterioration of a person’s state of health. Medical device software may fall under class I only if it is not intended to provide information used to make decisions for diagnostic or therapeutic purposes. For class I devices, no notified body is involved in the declaration of conformity.

These classification criteria are very stringent and prompted formation of the Medical Device Coordination Group (MDCG), an official working group serving to advise the European Commission regarding medical devices [8]. The group recently released a guidance document that further elaborates rule 11 [6]. The guidance provides examples on how to classify software, and the authors seem to interpret rule 11 less strictly than an original reading of the MDR would suggest. The document suggests for instance that software “intended to rank therapeutic suggestions for a health care professional based on patient history, imaging test results, and patient characteristics…, should be classified as class IIa. ”. If one merely reads rule 11, one could conclude that these types of software would fall within class III, since an error might cause death. This guidance is not legally binding [6] but notified bodies might consider it when making their decision.

Conformity Assessment Routes

Fulfilling the “General safety and performance requirements” described in Annex I of the MDR is the most crucial step on the long road to marketing a medical device. To prove compliance with these requirements, manufacturers must follow one of the conformity assessment procedures described in the MDR appendices. Typically, manufacturers apply harmonized standards, and in the future also common specifications, to prove compliance with these requirements (Articles 8-9 and Annex II [4C] MDR). Examples of general safety and performance requirements include risk management, software lifecycle processes, software verification and validation, and usability (Annex I MDR). A complete list of requirements can be found in Annex I of the MDR.

Conformity assessment is a process demonstrating whether the “general safety and performance requirements” of the MDR (Annex I MDR) have been fulfilled. Once the conformity of the medical device with the requirements has been proven, the manufacturer may declare the conformity, CE mark and market the product within the EU/EEA (Articles 19-20 MDR). Depending on the risk class, the MDR describes three different paths of conformity assessment in accordance with Annexes IX, X and XI (Article 56 MDR). Besides the four main risk classes (I, IIa, IIb, III), the MDR defines three sub-classes for risk class I, which are devices with measuring function (Im), sterile devices (Is) and reusable surgical instruments (Ir). For classes Im, Is, Ir, IIa, IIb and III, a notified body must be involved in the process of conformity assessment. This is not required for all remaining devices falling within class l (Articles 52-53 MDR). For risk class III, an additional expert panel will scrutinize the clinical evaluation and assess whether the clinical data is sufficient to provide confidence in the safety and performance of the device (Annex IX MDR). For medical device software, the development process cannot be ignored since it is difficult to find errors by simply testing the finished product, which would be the case in the Annex XI (Part B) conformity assessment. In contrast, the procedure described in Annex IX includes an assessment of the QMS and the technical documentation by a notified body. The product of a successful assessment is an Annex IX certificate and an EU QMS certificate for the manufacturer, who can then declare conformity of the medical device with the requirements set by the MDR (Articles 19 and 56 MDR).

Technical Documentation

According to the “general obligations of manufacturers” (Article 10[4] MDR), technical documentation must be compiled and kept up to date to enable assessment of compliance with the safety and performance requirements set by the MDR. Annex II of the MDR lists in detail what is required in the documentation, including documentation of a fully implemented risk management system, benefit-risk analysis, clinical evaluation report, software life cycle file, usability file, and many other requirements.

Clinical Evaluation

The supporting documentation for the CE declaration must include a clinical evaluation. This is an evaluation of side effects and the acceptability of the benefit-risk ratio based on clinical data (Article 61 MDR). Conducting a proper clinical evaluation will demonstrate (1) which clinical data are necessary; (2) which clinical data can be adequately supplemented by methods other than clinical investigations, such as published literature, prior clinical investigations, clinical experience, or by using suitable clinical data from equivalent devices; and (3) which clinical data remain to be delivered by clinical investigations (Article 61 MDR). Clinical investigations within the MDR are what many people would refer to as “clinical trials” and are defined as “systematic investigation involving one or more human subjects, undertaken to assess the safety or performance of a device” (Article 2[45] MDR). It is one of the methods to obtain clinical data supporting treatment efficacy and to confirm clinical benefit (Article 62 MDR). If sufficient clinical data to perform a clinical evaluation can be retrieved from the literature or other sources, the manufacturer can proceed without a clinical investigation. The clinical evaluation must be updated frequently with data from post-market surveillance. New data, as well as considerations for new or changed intended purposes, require an updated clinical evaluation and may indicate the necessity for additional clinical investigations (Article 61 MDR).

The clinical evaluation must be planned, conducted, and documented. The clinical evaluation, its results, and the clinical evidence derived from it must be documented in a clinical evaluation report and included as part of the technical documentation (Annexes II and XIV MDR[9]). Furthermore, a clinical evaluation plan must be elaborated and documented in the technical documentation (Annexes II and XIV MDR).

Regulatory Requirements for the Clinical Evaluation

App developers are advised to follow the regulations and guidelines listed in Textbox 3.